Hide metadata

dc.date.accessioned2016-06-07T14:00:15Z
dc.date.available2016-06-07T14:00:15Z
dc.date.created2016-02-05T08:41:18Z
dc.date.issued2016
dc.identifier.citationMathiassen, Kim Fjellin, Jørgen Enger Glette, Kyrre Harald Hol, Per Kristian Elle, Ole Jakob . An Ultrasound Robotic System Using the Commercial Robot UR5. Frontiers in Robotics and AI. 2016
dc.identifier.urihttp://hdl.handle.net/10852/50439
dc.description.abstractThe use of robots in health care has increased dramatically over the last decade. One area of research has been to use robots to conduct ultrasound examinations, either controlled by a physician or autonomously. This paper examines the possibility of using the commercial robot UR5 from Universal Robots to make a tele-operated robotic ultrasound system. Physicians diagnosing patients using ultrasound probes are prone to repetitive strain injuries, as they are required to hold the probe in uncomfortable positions and exert significant static force. The main application for the system is to relieve the physician of this strain by letting the them control a robot that holds the probe. A set of requirements for the system is derived from the state-of-the-art systems found in the research literature. The system is developed through a low-level interface for the robot, effectively building a new software framework for controlling it. Compliance force control and forward flow haptic control of the robot was implemented. Experiments are conducted to quantify the performance of the two control schemes. The force control is estimated to have a bandwidth of 16.6 Hz, while the haptic control is estimated to have a bandwidth of 65.4 Hz for the position control of the slave and 13.4 Hz for the force control of the master. Overall, the system meets the derived requirements and the main conclusion is that it is feasible to use the UR5 robot for robotic ultrasound applications.en_US
dc.languageEN
dc.language.isoenen_US
dc.publisherFrontiers Media S.A.
dc.relation.ispartofMathiassen, Kim (2017) A semi-autonomous robotic system for needle tracking and visual servoing using 2D medical ultrasound. Doctoral thesis. http://hdl.handle.net/10852/59088
dc.relation.urihttp://hdl.handle.net/10852/59088
dc.rightsAttribution 4.0 International
dc.rights.urihttps://creativecommons.org/licenses/by/4.0/
dc.titleAn Ultrasound Robotic System Using the Commercial Robot UR5en_US
dc.typeJournal articleen_US
dc.creator.authorMathiassen, Kim
dc.creator.authorFjellin, Jørgen Enger
dc.creator.authorGlette, Kyrre Harald
dc.creator.authorHol, Per Kristian
dc.creator.authorElle, Ole Jakob
cristin.unitcode185,53,18,17
cristin.unitnameIntervensjonssenteret
cristin.ispublishedtrue
cristin.fulltextoriginal
cristin.qualitycode1
dc.identifier.cristin1332470
dc.identifier.bibliographiccitationinfo:ofi/fmt:kev:mtx:ctx&ctx_ver=Z39.88-2004&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.jtitle=Frontiers in Robotics and AI&rft.volume=&rft.spage=&rft.date=2016
dc.identifier.jtitleFrontiers in Robotics and AI
dc.identifier.volume3
dc.identifier.doihttp://dx.doi.org/10.3389/frobt.2016.00001
dc.identifier.urnURN:NBN:no-54032
dc.type.documentTidsskriftartikkelen_US
dc.type.peerreviewedPeer reviewed
dc.source.issn2296-9144
dc.identifier.fulltextFulltext https://www.duo.uio.no/bitstream/handle/10852/50439/1/frobt-03-00001.pdf
dc.type.versionPublishedVersion
cristin.articleid1


Files in this item

Appears in the following Collection

Hide metadata

Attribution 4.0 International
This item's license is: Attribution 4.0 International