dc.identifier.citation | Henriksen, Hege Berg Henriksen, Christine Sequeira de Sousa, Ana Rita Alavi, Dena Helene Augestad, Elin Marie Rising, Russell Dörner, Rebecca Hägele, Franziska A. Bosy-Westphal, Anja Blomhoff, Rune Ulven, Stine Marie Olsen, Thomas . Validity and reproducibility of a whole-room indirect calorimeter for the estimation of VO2, VCO2, and resting metabolic rate. Physiological Reports. 2023, 11(7) | |
dc.identifier.uri | http://hdl.handle.net/10852/107924 | |
dc.description.abstract | Abstract Whole‐room indirect calorimeters (WRICs) provide accurate instruments for the measurement of respiratory exchange, energy expenditure, and macronutrient oxidation. Here, we aimed to determine the validity and reproducibility of a 7500 L WRIC for the measurement of ventilation rates and resting metabolic rate (RMR). Technical validation was performed with propane combustion tests ( n = 10) whereas biological reproducibility was tested in healthy subjects (13 women, 6 men, mean ± SD age 39.6 ± 15.3) in two 60 min measurements separated by 24 h. Subjects followed a run‐in protocol prior to measurements. The coefficient of variation (CV) and intraclass correlation coefficient (ICC) were calculated for ventilation rates of O2 (VO2), CO2 (VCO2), the respiratory quotient (RQ; VCO2/VO2), and RMR. Technical validation showed good validity with CVs ranging from 0.67% for VO 2 to 1.00% for energy expenditure. For biological reproducibility, CVs were 2.89% for VO 2 ; 2.67% for VCO 2 ; 1.95% for RQ; and 2.68% for RMR. With the exception of RQ (74%), ICCs were excellent for VO 2 (94%), VCO 2 (96%) and RMR (95%). Excluding participants that deviated from the run‐in protocol did not alter results. In conclusion, the 7500 L WRIC is technically valid and reproducible for ventilation rates and RMR. | |
dc.language | EN | |
dc.rights | Attribution 4.0 International | |
dc.rights.uri | https://creativecommons.org/licenses/by/4.0/ | |
dc.title | Validity and reproducibility of a whole-room indirect calorimeter for the estimation of VO2, VCO2, and resting metabolic rate | |
dc.title.alternative | ENEngelskEnglishValidity and reproducibility of a whole-room indirect calorimeter for the estimation of VO2, VCO2, and resting metabolic rate | |
dc.type | Journal article | |
dc.creator.author | Henriksen, Hege Berg | |
dc.creator.author | Henriksen, Christine | |
dc.creator.author | Sequeira de Sousa, Ana Rita | |
dc.creator.author | Alavi, Dena Helene | |
dc.creator.author | Augestad, Elin Marie | |
dc.creator.author | Rising, Russell | |
dc.creator.author | Dörner, Rebecca | |
dc.creator.author | Hägele, Franziska A. | |
dc.creator.author | Bosy-Westphal, Anja | |
dc.creator.author | Blomhoff, Rune | |
dc.creator.author | Ulven, Stine Marie | |
dc.creator.author | Olsen, Thomas | |
cristin.unitcode | 185,51,13,0 | |
cristin.unitname | Avdeling for ernæringsvitenskap | |
cristin.ispublished | true | |
cristin.fulltext | original | |
cristin.qualitycode | 1 | |
dc.identifier.cristin | 2139735 | |
dc.identifier.bibliographiccitation | info:ofi/fmt:kev:mtx:ctx&ctx_ver=Z39.88-2004&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.jtitle=Physiological Reports&rft.volume=11&rft.spage=&rft.date=2023 | |
dc.identifier.jtitle | Physiological Reports | |
dc.identifier.volume | 11 | |
dc.identifier.issue | 7 | |
dc.identifier.doi | https://doi.org/10.14814/phy2.15658 | |
dc.type.document | Tidsskriftartikkel | |
dc.type.peerreviewed | Peer reviewed | |
dc.source.issn | 2051-817X | |
dc.type.version | PublishedVersion | |
cristin.articleid | e15658 | |